Fluorescent HPLC assay for 20-HETE and other P-450 metabolites of arachidonic acid.
This study describes a fluorescent HPLC assay for measuring 20-hydroxyeicosatetraenoic acid (20-HETE) and other cytochrome P-450 metabolites of arachidonic acid in urine, tissue, and interstitial fluid. An internal standard, 20-hydroxyeicosa-6(Z),15(Z)-dienoic acid, was added to samples, and the lipids were extracted and labeled with 2-(2,3-naphthalimino)ethyl trifluoromethanesulfonate. P-450 metabolites were separated on a C18 reverse-phase HPLC column. Coelution and gas chromatography-mass spectrometry studies confirmed the identity of the 20-HETE peak. The 20-HETE peak can be separated from those for dihydroxyeicosatrienoic acids, other HETEs, and epoxyeicosatrienoic acids. Known amounts of 20-HETE were used to generate a standard curve (range 1-10 ng, r(2) = 0.98). Recovery of 20-HETE from urine averaged 95%, and the intra-assay variation was <5%. Levels of 20-HETE were measured in 100 microliter of urine and renal interstitial fluid or 0.1 mg of renal tissue. The assay was evaluated by studying the effects of 1-aminobenzotriazole (ABT) on the excretion of 20-HETE in rats. ABT reduced excretion of 20-HETE by >65% and inhibited the formation of 20-HETE by renal microsomes. The availability of this assay should facilitate work in this field.[1]References
- Fluorescent HPLC assay for 20-HETE and other P-450 metabolites of arachidonic acid. Maier, K.G., Henderson, L., Narayanan, J., Alonso-Galicia, M., Falck, J.R., Roman, R.J. Am. J. Physiol. Heart Circ. Physiol. (2000) [Pubmed]
Annotations and hyperlinks in this abstract are from individual authors of WikiGenes or automatically generated by the WikiGenes Data Mining Engine. The abstract is from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.About WikiGenesOpen Access LicencePrivacy PolicyTerms of Useapsburg