Non-extraction HPLC method for simultaneous measurement of dyphylline and doxofylline in serum.
A direct injection HPLC method for the simultaneous measurement of dyphylline and doxofylline in serum is reported. Chromatography is carried out with a "Pinkerton" internal surface reversed-phase column and phosphate buffer (0.1 mol/L, pH 6.8) as mobile phase. Absorbance at 275 nm is monitored. Only 10 microL of serum is required to detect less than 1 mg of each drug per liter in biological samples. The standard curve for the method is linear over the range 6-100 mg/L, and precision is acceptable for both drugs, with CVs of 2% and 1.5% for respective concentrations of 6.2 and 25 mg/L. None of the 76 drugs tested for interference affected the measurement of either drug. Four samples can be injected per hour. This method provides a fast, accurate way to monitor two interesting therapeutic agents used in chronic obstructive airway diseases.[1]References
- Non-extraction HPLC method for simultaneous measurement of dyphylline and doxofylline in serum. Tagliaro, F., Dorizzi, R., Frigerio, A., Marigo, M. Clin. Chem. (1990) [Pubmed]
Annotations and hyperlinks in this abstract are from individual authors of WikiGenes or automatically generated by the WikiGenes Data Mining Engine. The abstract is from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.About WikiGenesOpen Access LicencePrivacy PolicyTerms of Useapsburg