Analysis and comparison of venous air embolism detection methods.
Changes in precordial ultrasonic Doppler patterns in dogs were compared with changes of end-tidal carbon dioxide (EtCO2) and pulmonary arterial pressure (PAP). Possible mechanisms of pulmonary arterial hypertension after air embolism in dogs are discussed. After the injection of air, an immediate change in the Doppler ultrasonic pattern was detected, including a shift of the base line. The duration of change of Doppler patterns varied depending upon the length of time that trapped air remained in the right heart. There was a delay of 15 to 30 seconds before the EtCO2 and PAP responded. In a group of dogs pretreated with 0.3 mg of propranolol per kg, there was a significant decrease in the PAP response due to induced air embolism compared to the control group (P < 0.05). We found that the Doppler ultrasonic device was the most sensitive for detecting venous air embolism at a level as low as 0.05 ml of air per kg compared to 0.25 ml/kg for the PAP and EtCO2 responses.[1]References
- Analysis and comparison of venous air embolism detection methods. Chang, J.L., Albin, M.S., Bunegin, L., Hung, T.K. Neurosurgery (1980) [Pubmed]
Annotations and hyperlinks in this abstract are from individual authors of WikiGenes or automatically generated by the WikiGenes Data Mining Engine. The abstract is from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.About WikiGenesOpen Access LicencePrivacy PolicyTerms of Useapsburg